Dose-effect calibration curve for high X-ray doses using the Calyculin-A chromosome premature condensation assay

Título traducido de la contribución: Curva de calibración dosis-efecto para altas dosis de rayos X utilizando el ensayo de condensación prematura de cromosomas con Caliculina-A
  • F. A. Chaves-Campos
  • , F. Ortíz-Morales
  • , R. Mejías-Gamboa
  • , J. E. González-Mesa
  • , O. García-Lima
  • , M. P. Rodríguez-Valerio
  • , W. Vargas-Segura
  • , A. Cordero-Ramírez
  • University of Costa Rica
  • School of Health Technologies
  • Robotic Radiosurgery Center
  • Latin American Biological Dosimetry Network (LBDNet)
  • Centro de Protección e Higiene de las Radiaciones (CPHR)
  • Hospital México
  • Radiotherapy Department

Producción científica: Contribución a una revista › Artículo › revisión exhaustiva

4 Citas (Scopus)

Resumen

<jats:p>Purpose: This article shows the results of the Cytogenetics Laboratory of the Health Research Institute (INISA) to develop a dose-effect calibration curve with the Calyculin-A chemical induction premature condensation assay to estimate high doses of X-ray exposure. Methods: to create the calibration curve, peripheral blood samples from two participants (one female and one male) were exposed to X-rays at six different dose points ranging from 0 to 17.5 Gy in vitro. The irradiated blood was cultured for 48 hours according to international protocols, and the resulting chromosome rings were recorded. We used BioDoseTools software to calculate the coefficients for the calibration curve. Results: The coefficients of the curve are α: 0.028±0.001 and C: 0.001±0.001. These coefficients have similar values to those reported internationally. The curve was validated by calculating an unknown dose exposed to 6 Gy; the estimated dose was 5.651 ± 0.636 Gy, with no statistically significant differences between the dose delivered and the estimated dose. Conclusions: The INISA Biological Dosimetry Service can use the curve obtained to assess absorbed doses in cases of suspected overexposure to high X-ray doses.</jats:p>
Título traducido de la contribuciónCurva de calibración dosis-efecto para altas dosis de rayos X utilizando el ensayo de condensación prematura de cromosomas con Caliculina-A
Idioma originalInglés
Número de artículoe2422
PublicaciónBrazilian Journal of Radiation Sciences
Volumen12
N.º2
DOI
EstadoPublicada - 24 may 2024

Huella

Profundice en los temas de investigación de 'Curva de calibración dosis-efecto para altas dosis de rayos X utilizando el ensayo de condensación prematura de cromosomas con Caliculina-A'. En conjunto forman una huella única.