TY - GEN
T1 - Evaluación experimental del tiempo de respuesta y recuperación en contadores Geiger-Muller
AU - Vargas-Segura, Walter
AU - Jiménez-Carballo, Carlos A.
N1 - Publisher Copyright:
© 2026 Latin American and Caribbean Consortium of Engineering Institutions. All rights reserved.
PY - 2026
Y1 - 2026
N2 - This work experimentally evaluates the dynamic behavior of digital and analog Geiger-Müller (GM counters under operational conditions in a facility housing ionizing radiation sources. Although these instruments are widely used in radiation protection, their temporal response to abrupt changes in dose is not always considered in practice. For this study, quantitative criteria were defined to determine the response time (exceeding 90 % of the maximum recorded dose rate) and the recovery time (returning withing 10 % of the reference value). Measurements were obtained through video analysis with a temporal resolution of 1/30 s, and five repetitions were performed for each condition. Time uncertainty was estimated as Type A in accordance with the GUM, and ANOVA and Tukey tests (α = 0, 05) were applied to evaluate differences among indication channels (numerical, audible, and visual). The results show that the characteristic times, on the order of seconds, depend significantly on the instrument type and output channel, being greater for numerical indicators than for alarms. The observed behavior is dominated by signal processing electronics rather than detector dead time, indicating that quantitative reading is not strictly instantaneous in scenarios involving rapid dose rate variations.
AB - This work experimentally evaluates the dynamic behavior of digital and analog Geiger-Müller (GM counters under operational conditions in a facility housing ionizing radiation sources. Although these instruments are widely used in radiation protection, their temporal response to abrupt changes in dose is not always considered in practice. For this study, quantitative criteria were defined to determine the response time (exceeding 90 % of the maximum recorded dose rate) and the recovery time (returning withing 10 % of the reference value). Measurements were obtained through video analysis with a temporal resolution of 1/30 s, and five repetitions were performed for each condition. Time uncertainty was estimated as Type A in accordance with the GUM, and ANOVA and Tukey tests (α = 0, 05) were applied to evaluate differences among indication channels (numerical, audible, and visual). The results show that the characteristic times, on the order of seconds, depend significantly on the instrument type and output channel, being greater for numerical indicators than for alarms. The observed behavior is dominated by signal processing electronics rather than detector dead time, indicating that quantitative reading is not strictly instantaneous in scenarios involving rapid dose rate variations.
KW - Geiger-Müller counter
KW - nuclear instrumentation
KW - radiation protection
KW - recovery time
KW - response time
UR - https://www.scopus.com/pages/publications/105048855266
U2 - 10.18687/LACCEI2026.1.1.2386
DO - 10.18687/LACCEI2026.1.1.2386
M3 - Contribución a la conferencia
AN - SCOPUS:105048855266
T3 - Proceedings of the LACCEI international Multi-conference for Engineering, Education and Technology
BT - Proceedings of the 24th LACCEI International Multi-Conference for Engineering, Education and Technology - Engineering without Borders
A2 - Petrie, Maria M. Larrondo
A2 - Texier, Jose
A2 - Matta, Rodolfo Andres Rivas
PB - Latin American and Caribbean Consortium of Engineering Institutions
T2 - 24th LACCEI International Multi-Conference for Engineering, Education and Technology, LACCEI 2026
Y2 - 16 June 2026 through 18 June 2026
ER -